If you want your monitor to communicate with the meter, the following conditions must apply:
1 The meter option must be turned on and programmed. Refer to the instructions in this section.
2 Your monitor must be within 4 feet (1.2 meters) of your meter to receive the blood glucose reading.
3 The monitor cannot have a LOW BATTERY alert condition.
4 The blood glucose measurement from the meter will appear as the default blood glucose value on the
ENTER BG screen. The monitor will not display a reading that is older than 12 minutes on the ENTER
BG screen.
5 Do not use the RF meter to send your blood glucose readings to the monitor while onboard aircraft.
Manually enter your blood glucose.
Add, delete, review meter IDs
Select Meter Options in the Utilities Menu screen when programming your meter.
If you are not sure that your meter ID is entered in your monitor check the REVIEW METER ID screen.
You have to turn on the meter option to add, delete, or review the meter ID(s) programmed in your
monitor.
1 Go to the METER OPTION screen:
Main > Utilities > Meter Options > Meter Option
2 Select On, then press ACT. The METER ID MENU appears.
3 Add, delete or review your meter ID(s) as desired.
Adding meter IDs
a. Select Add ID, then press ACT.
b. Use the up and down arrow buttons to enter each of the six ID characters. Press ACT after each
entry.
c. After you set the last character of the ID, the screen will return to the METER ID MENU.
Deleting meter IDs
a. Select Delete ID, then press ACT.
b. Select the meter ID that you want to delete and press ACT.
c. The selected ID is now deleted.
Meter option44