Follow these steps when you get an alarm or alert:
To view the alarm/alert:
1 From the HOME screen, review the alarm/alert by pressing any button to see the message.
2 Read the entire alarm or alert message. The message includes instructions on how to fix the alarm
or alert condition. Press
to scroll if the message is bigger than your screen.
3 Press ESC to clear the alarm/alert, then press ACT after you read the alarm/alert instructions. The
HOME screen appears.
4 Follow the instructions in the alarm/alert message to fix the alarm/alert condition.
5Check your settings (for example, time and date and Glucose Limits) to make sure that they are
correct.
Sensor alerts
You may encounter the alerts listed below while using the system. These alerts include information about
how to resolve the condition.
WEAK SIGNAL
Reason: Occurs when the monitor does not receive data from the transmitter for a pre-defined period of
time (as set in Missed Data).
Action: Move the monitor closer to the transmitter or move the transmitter and the monitor to a new
location on your body.
Troubleshooting and alarms/alerts84