872 Extension Module IC Pump – 2.872.0010 Manual 8.872.
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 872 Extension Module IC Pump – 2.872.0010 Manual 8.872.8005EN 08.
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above. Documentation in additional languages can be found on http://products.metrohm.com under Literature/Technical documentation.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1.1 1 Instrument description ......................................................... 1 1.2 About the documentation ................................................... 2 1.2.1 Content and scope .................................................................. 2 1.2.2 Symbols and conventions ........................................................ 2 1.3 Safety instructions ........................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 4.8 Pulsation damper ............................................................... 35 5 Start-up 37 6 Operation and maintenance 38 6.1 General information ........................................................... 38 6.1.1 Care ...................................................................................... 38 6.1.2 Maintenance by Metrohm Service .......................................... 38 6.1.3 Operation ...........................................
■■■■■■■■■■■■■■■■■■■■■■ 10 Accessories Index 872 Extension Module IC Pump Table of contents 63 10.1 Scope of delivery ................................................................ 63 10.2 Optional accessories ...........................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Figure 25 Figure 26 Figure 27 Figure 28 Figure 29 Figure 30 Figure 31 Figure 32 Figure 33 Figure 34 VI ■■■■■■■■ Front 872 Extension Module IC Pump ................................................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 Instrument description Existing 850 Professional IC instruments can be expanded to include additional functions by means of 872 Extension Modules. Every 850 Professional IC instrument can be supplemented with up to 3 extension modules. The 872 Extension Module – IC Pump permits the installation of an additional IC high pressure pump «iPump» in a Professional IC System.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 About the documentation Pulsation damper The pulsation damper protects the separation column from damage caused by pressure fluctuations when switching the injection valve, and reduces interfering pulsations during highly sensitive measurements. 1.2 About the documentation Caution Please read through this documentation carefully before putting the instrument into operation.
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction Warning This symbol draws attention to a possible hazard due to electrical current. Warning This symbol draws attention to a possible hazard due to heat or hot instrument parts. Warning This symbol draws attention to a possible biological hazard. Caution This symbol draws attention to a possible damage of instruments or instrument parts. Note This symbol marks additional information and tips. 1.3 Safety instructions 1.3.
■■■■■■■■■■■■■■■■■■■■■■ 1.3 Safety instructions Warning Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched. There are no parts inside the housing which can be serviced or replaced by the user. Mains voltage Warning An incorrect mains voltage can damage the instrument. Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
■■■■■■■■■■■■■■■■■■■■■■ 1.3.4 1 Introduction Flammable solvents and chemicals Warning All relevant safety measures are to be observed when working with flammable solvents and chemicals. ■ ■ ■ ■ 1.3.5 Set up the instrument in a well-ventilated location (e.g. laboratory flue). Keep all sources of flame far from the workplace. Clean up spilled fluids and solids immediately. Follow the safety instructions of the chemical manufacturer.
■■■■■■■■■■■■■■■■■■■■■■ 2.1 Front 2 Overview of the instrument 2.1 Front 5 1 7 3 Figure 1 2 6 4 Front 872 Extension Module IC Pump 1 Eluent degasser See Chapter 4.5 2 Coupling 6.2744.230 For connecting the eluent aspiration tubing. 3 High pressure pump See Chapter 4.6 4 Purge valve See Chapter 4.6 5 Inline filter See Chapter 4.7 6 Pulsation damper See Chapter 4.
■■■■■■■■■■■■■■■■■■■■■■ 2.2 2 Overview of the instrument Rear 1 to Vacuum Transport security screws Out In Made by Metrohm Herisau Switzerland 2 Figure 2 4 3 Rear 872 Extension Module IC Pump 1 Transport locking screws 2 Connector In To connect the extension module to the IC instrument or to a previous extension module. 3 Connector Out To connect an additional extension module.
■■■■■■■■■■■■■■■■■■■■■■ 3.1 General 3 Assembly 3.1 General The extension modules are fitted directly to the 850 Professional IC instrument and connected with it via 6.2156.060 connection cable. Extension modules have no power supply of their own, but rather draw the electricity they require from the instrument with which they are connected. Up to three extension modules can be connected to an 850 Professional IC instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly A Figure 3 C B Setup versions A Extension module above the IC instrument Between the 850 Professional IC and the bottle holder. C Extension module next to the IC instrument With its own base plate and its own bottle holder next to the 850 bottle holder. B Extension module on the below the IC instrument Between the base plate and the 850 Profession IC. Position the extension module in such a way that the capillary connections can be kept as short as possible.
■■■■■■■■■■■■■■■■■■■■■■ 3.2 Mounting the extension module onto the IC instrument 4 Dismounting the bottle holder 1 1 2 2 Figure 4 1 Dismounting the bottle holder 2 Cover stoppers ■ ■ ■ Cylinder screws Remove covering stoppers (4-1). Loosen the cylinder screws with a 6.2621.100 3 mm hexagon key. Remove the bottle holder 5 Attaching the extension module(s) Place the extension module(s) on the IC instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly 6 Mounting the bottle holder 1 1 2 2 Figure 5 1 Mounting the bottle holder 2 Cover stoppers ■ ■ ■ Cylinder screws Attach the bottle holder on the extension module. Tighten the cylinder screws (4-2) with an 6.2621.100 3 mm hexagon key Insert covering stoppers (4-1). 7 Connecting the extension module ■ ■ Plug a 6.2156.060 cable into the connector In of the extension module and screw it tight.
3.3 Mounting the extension module below the IC instrument ■■■■■■■■■■■■■■■■■■■■■■ 9 Mounting the drainage tubing Reconnect the drainage tubing to the drainage tubing connector of the bottle holder. Possibly, a longer section of silicone tubing 6.186.020 must be cut to fit and mounted (see also the manual for the IC instrument). 3.3 Mounting the extension module below the IC instrument 1 Switching off the IC instrument Switch the IC instrument off and disconnect the mains cable.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly 2 1 2 Figure 6 1 Removing the base tray 2 Base tray Cylinder screws With washer. 6 Mounting the base tray ■ ■ ■ 872 Extension Module IC Pump Tilt the extension module sideways and lay it down flat. Attach base tray. Slide the washers onto the cylinder screws (7-2) and tighten these with a 3 mm hexagon key (6.2621.100).
■■■■■■■■■■■■■■■■■■■■■■ 3.3 Mounting the extension module below the IC instrument 2 1 2 Figure 7 1 Mounting the base tray 2 Base tray ■ ■ ■ Cylinder screws With washer. Set up the extension module. Optional: Set up further extension modules. Set up the IC instrument. 7 Connecting the extension module ■ ■ Plug a 6.2156.060 cable into the connector In of the extension module and screw it tight.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly 9 Inserting the detector(s) again and connecting it (them) See manual for the IC instrument. 10 Restoring the loosened connections ■ ■ ■ ■ ■ 3.4 Reconnect the drainage tubings, Possibly, a longer section of silicone tubing 6.186.020 must be cut to fit and mounted (see also the manual for the IC instrument). Connect the leak sensor (see manual for the IC instrument), Connect the USB cable, Connect the MSB cable, Plug in the mains cable.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Setting up the extension module next to the IC instrument 2 1 2 Figure 8 1 2 Base tray ■ ■ 16 ■■■■■■■■ Mounting the base tray Cylinder screws With washer. Set up the extension module. Optional: Set up further extension modules.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly 3 Mounting the bottle holder 1 1 2 2 Figure 9 1 Mounting the bottle holder 2 Cover stoppers ■ ■ ■ Cylinder screws Place the bottle holder on the extension module. Tighten the cylinder screws (9-2) with an 6.2621.100 3 mm hexagon key. Insert covering stoppers (9-1). 4 Connecting the extension module ■ ■ Plug a 6.2156.060 cable into the connector In of the extension module and screw it tight.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Setting up the extension module next to the IC instrument 6 Connecting the leak sensor ■ ■ ■ Plug the 6.2103.170 adapter into the leak sensor connector of the IC instrument. Connect the leak sensor cable of the IC instrument to the adapter. Connect the leak sensor cable of the extension module to the adapter. 7 Connecting the drainage tubings 1 2 4 3 5 Figure 10 1 Connecting the drainage tubings Drainage tubing connector For draining escaped liquid from the bottle holder.
■■■■■■■■■■■■■■■■■■■■■■ 3 Assembly 3 Drainage tubing connection For supplying escaped liquid through the connected drainage tubing to the leak sensor. 5 Drainage tubing Section of the 6.1816.020 silicon tubing. Guides escaped liquid into a waste container. ■ ■ ■ 3.5 4 Drainage tubing connector For draining escaped liquid from the base tray through the connected drainage tubing.
■■■■■■■■■■■■■■■■■■■■■■ 3.5 Connecting the vacuum pump(s) 2 1 4 “to Vacuum” 872 Extension Module 1 “to Vacuum” 872 Extension Module 2 Figure 11 3 “Vacuum” 850 Professional IC Connecting the vacuum pump 1 T connector 6.1808.060 For connecting several FEP tubings. 2 FEP tubing 6.1805.050 Connect to the connector to Vacuum on the first 872 Extension Module. 3 FEP tubing 6.1805.050 Connect to the connector to Vacuum on the second 872 Extension Module. 4 FEP tubing 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation 4 Installation 4.1 About this chapter The Installation chapter contains ■ ■ ■ ■ 4.2 this overview a brief set of instructions for the installation of the 872 Extension Module IC Pump (see Chapter 4.2, page 21). At each step you will find cross-references to detailed installation instructions for individual components, should you require such aids. two installation diagrams (see Chapter 4.3, page 22), which show two possible applications of the instrument.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 Installation diagram ■ Connect the other end of the 6.1831.100 connection capillary (12-2)/(13-2) either with – a 6.2744.070 pressure screw to the T distributor (12-7) of the gradient mixing coil (gradient system) or – a 6.2744.014 pressure screw to the injection valve of the 850 Professional IC instrument (isocratic system). 2 Putting extension module into operation See Chapter 5, Page 37 ■ ■ 4.
■■■■■■■■■■■■■■■■■■■■■■ High pressure gradient system 4 Installation 6 5 6 4 1 3 7 2 3 Figure 12 Installation diagram – gradient 1 Eluent aspiration tubing 6.1834.080 Connection eluent bottle – eluent degasser 2 Connection capillary 6.1831.100 Connection pulsation damper – gradient mixing coil 3 PEEK pressure screw, short 6.2744.070 4 Connection capillary 6.1831.100 Connection to another high pressure pump 5 Gradient mixing coil 6.2758.000 6 PEEK pressure screw 6.2744.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 Eluent Isocratic system 1 2 3 Figure 13 Installation diagram – isocratic 1 Eluent aspiration tubing 6.1834.080 Connection eluent bottle – eluent degasser 3 PEEK pressure screw, short 6.2744.070 2 Connection capillary 6.1831.100 Connection pulsation damper – IC instrument The following chapters describe the individual installation steps in detail. 4.4 Eluent 4.4.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Assembling eluent aspiration tubing 1 Guide the free end of the eluent aspiration tubing (14-1) out of the instrument through a suitable capillary feed-through. 2 Installing the 6.1602.160 eluent bottle attachment Slide tubing nipple (14-2) and O-ring (14-3) onto the eluent aspiration tubing (14-1). Push eluent aspiration tubing (14-1) through the bottle attachment (14-4) and screw tight.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 Eluent 4 Installing tubing weighting and aspiration filter 1 3 2 Figure 16 4 5 Installing tubing weighting and aspiration filter 1 Eluent aspiration tubing 6.1834.080 2 Eluent bottle attachment 6.1602.160 3 Tubing weighting From accessory set 6.2744.210. 4 Clamping screw From accessory set 6.2744.210. 5 Aspiration filter 6.2821.090 With filter holder from accessory set 6.2744.210.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Fasten the bottle attachment (16-2) on the eluent bottle (18-10). The aspiration filter (18-6) must rest on the base of the eluent bottle. ■ 6 Mounting the adsorber tube Note In the case of alkaline eluents and eluents with lower buffer capacity, the eluent bottle must be equipped with a CO2 adsorber (18-4). First, place a piece of wadding (18-3), then the CO2 adsorber (18-4) in the large opening of the adsorber tube (18-2) and close with the plastic cover.
■■■■■■■■■■■■■■■■■■■■■■ 4.5 Eluent degasser 5 Eluent 6 Aspiration filter 6.2821.090 7 Filter holder From accessory set 6.2744.210. 8 Clamping screw From accessory set 6.2744.210. 9 Tubing weighting From accessory set 6.2744.210. 10 Eluent bottle 6.1608.070 11 Bottle attachment 6.1602.160 12 SGJ clip 6.2023.020 13 Tubing nipple 14 Thread stopper 4.5 Eluent degasser Gas bubbles in the eluent lead to an unstable baseline, as high pressure pumps can transport liquids, but not gases.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation 3 Tubing flare With tubing nipple. 4 Clamping screw 5 Eluent aspiration tubing 6.1834.080 For aspirating the eluent. The clamping screw (19-4) is firmly mounted. 6 Connection capillary 6.1834.090 Connection from the eluent degasser to the high pressure pump (see Chapter 4.6, page 29). The clamping screw (19-4) is firmly mounted. 1 Caution The clamping screws (19-4) must be tightened carefully. Use the 6.2621.050 wrench for this. ■ ■ 2 ■ ■ ■ 4.
■■■■■■■■■■■■■■■■■■■■■■ 4.6 High pressure pump 2 13 1 2 3 4 12 5 2 6 2 11 2 7 10 2 8 9 Figure 20 2 Capillary connections high pressure pump/purge valve 1 Connection capillary PEEK capillary, connects main piston and auxiliary piston. 2 PEEK pressure screw, short 6.2744.070 3 Outlet valve holder 4 Pump head 6.2824.110 5 Fastening screws For fastening the pump head. 6 Inlet valve holder 7 Pump head input capillary PEEK capillary at the input of the pump head.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Note The eluent aspiration tubing is already installed in the newly delivered instrument. The following installation instructions need not be carried out at the time of initial installation. Connecting inlet to the high pressure pump 5 1 2 Figure 21 1 3 High pressure pump – Connect inlet Pressure screw For connecting the coupling (21-2) to the pump head input capillary (20-7). 2 Coupling 6.2744.
■■■■■■■■■■■■■■■■■■■■■■ 4.6 High pressure pump 4.6.2 Deaerating the high pressure pump The high pressure pump will only operate perfectly if the pump head contains no more air bubbles. Therefore it must be deaerated during initial start-up and after every change of eluent. Caution The high pressure pump must not be deaerated before the initial startup .
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Deaerating the high pressure pump The instrument must be connected to the PC and switched on to deaerate the high pressure pump. 5 4 5 6 3 7 2 5 1 Figure 22 Deaerating the high pressure pump 1 Syringe 10 mL 6.2816.020 For aspirating the eluent. 2 Luer connector On purging needle. 3 Purging needle 6.2816.040 4 Deaerating capillary 5 PEEK pressure screws, short 6.2744.
■■■■■■■■■■■■■■■■■■■■■■ 4.7 Inline filter 4 Setting the flow rate ■ ■ ■ Start MagIC Net (if not yet started). Ensure that the eluent aspiration tubing is immersed sufficiently in the eluent. Let the high pressure pump run. 5 Aspirating eluent ■ Aspirate with the syringe (22-1) until bubble-free eluent flows into the syringe. 6 Completing deaerating ■ ■ ■ ■ 4.7 Turn off high pressure pump. Close rotary knob (22-7). Remove syringe (22-1) from the Luer connector (22-2).
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Installing the inline filter Caution Observe the flow direction marked on the filter housing for the connection of the inline filter. 1 Figure 23 2 3 2 4 Connecting inline filter 1 Connection capillary Connects the purge valve with the inline filter 2 PEEK pressure screws, short 6.2744.070 3 Inline filter 6.2821.120 Protects against particles. 4 Connection capillary Connects the inline filter with the pulsation damper.
■■■■■■■■■■■■■■■■■■■■■■ 4.8 Pulsation damper The pulsation damper protects the separation column from damage caused by pressure fluctuations when switching the injection valve, and reduces interfering pulsations during highly sensitive measurements. In order to ensure these functionalities, it must be connected between the high pressure pump (see Chapter 4.6, page 29) and injection valve . The pulsation damper can be operated in both directions.
■■■■■■■■■■■■■■■■■■■■■■ 5 Start-up 5 Start-up The extension module is put into operation together with the IC instrument. Putting IC instrument into operation with extension modules 1 Start MagIC Net. 2 Connect IC instrument to PC and switch on. The extension module is recognized automatically by MagIC Net. 3 Deaerate the high pressure pump (see Chapter 4.6.2, page 32). Additional information can be found in the Start-up chapter in the manual for the IC instrument as well as in the MagIC Net online Help.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 General information 6 Operation and maintenance 6.1 General information 6.1.1 Care Warning The instrument housing must not be opened by untrained personnel. The instrument requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the service life of the sturdy mechanics and electronics.
■■■■■■■■■■■■■■■■■■■■■■ 6.1.4 6 Operation and maintenance Shutting down If the instrument is shut down for a longer period of time, the entire IC system must be rinsed as follows to rid it of salts in order to prevent eluent salts from forming crystals which may cause subsequent damage. rinse all capillaries and the Dosino with methanol/ultra pure water (1:4), rinse all pump tubings of the peristaltic pump with ultra pure water. ■ ■ 6.2 Door Caution The door is made of PMMA (polymethylmetacrylate).
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump 6.3.2 6.3.2.1 pH pH changes result in shifts in the dissociation equilibria and hence changes in the retention times. Organic solvents The addition of an organic solvent (e.g. methanol, acetone, acetonitrile) to aqueous eluents generally accelerates lipophilic ions. Operation Supply bottle The supply bottle with the eluent must be connected as indicated in Chapter 4.4.1, Page 24. This is above all important for eluents with volatile solvents (e.g.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance Salt crystals between the piston and seal cause abrasion particles which can find their way into the eluent. These lead to contaminated valves, a rise in pressure and in extreme cases scratched pistons. It is therefore essential to ensure that no precipitates can occur (see Chapter 6.3.2.3, page 40). Caution In order to spare the pump seals, the pump should not be operated dry.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump piston is on the left (viewed from the front), and the auxiliary piston is on the right. Cleaning/replacing the zirconium oxide piston Clean one piston after the other as follows: 1 Removing the piston cartridge from the pump head Loosen the piston cartridge with a wrench and unscrew from the pump head by hand.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance ■ ■ 1 Figure 26 2 3 Remove the spring retainer, spring and the inner plastic sleeve from the piston cartridge and lay by. Remove the backup ring from the pump head and lay to the other parts. 4 5 6 7 8 Components of the piston cartridge 1 Piston cartridge screw 2 Retaining washer 3 Zirconium oxide piston with piston shaft Order number: 6.2824.070. 4 Spring retainer 5 Spring Order number: 6.2824.060.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump Replacing the piston seal The 6.2617.010 special tool (see Figure 27, page 44) is necessary in order to remove the piston seal from the pump head. It consists of two parts: a tip for removing the old piston seal and a sleeve for inserting the new piston seal. 1 2 Figure 27 1 Tool for piston seal 6.2617.010 2 Pin Pin for removing the old piston seal. Sleeve Sleeve for inserting the new piston seal. Caution Screwing the 6.2617.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance 1 2 Figure 28 1 Removing the piston seal 2 Piston seal Tool for piston seal Pin of the tool. 2 Inserting the new piston seal into the tool Insert the new piston seal tightly by hand into the recess of the sleeve of the tool for the piston seal (27-2). The sealing springs must be visible from the outside. 1 2 Figure 29 1 Tool for piston seal 6.2617.010 Sleeve for inserting the new piston seal.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump 1 2 Figure 30 Inserting the piston seal into the pump head 4 Replacing the piston cartridge Screw the assembled piston cartridge back into the pump head and tighten, first by hand, then additionally by approx. 15° with a wrench. Cleaning the inlet valve and outlet valve 1 Removing valves ■ ■ 46 ■■■■■■■■ Unscrew the connection capillary for the auxiliary piston (20-1) from the outlet valve holder.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance 1 2 3 4 Figure 31 Removing valves 1 Outlet valve holder 2 Outlet valve Order number: 6.2824.160 3 Inlet valve Order number: 6.2824.170 4 Inlet valve holder 2 Cleaning undissected valve Clean contaminated or blocked valves initially without dismantling them completely. ■ ■ Rinse the valve in eluent flow and counterflow direction using a spray bottle filled with ultra pure water, RBS solution or acetone.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump ■ ■ Place the valve with the seal faced downwards above the recess in the holder. Push the valve components out of the valve housing using the needle of the tool. 1 2 3 4 Figure 32 Dismantling valve 1 Needle For pushing the valve components out of the valve housing. 2 Valve 3 Holder 4 Recess For collecting the valve components. The components of the valve are collected in the recess of the holder. Note The components of the valve are very small.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance 1 Figure 33 2 3 4 5 5 6 6 7 9 8 8 9 7 10 10 Components of the inlet valve and outlet valve 1 Inlet valve 6.2824.170 2 Outlet valve 6.2824.160 3 Inlet valve housing 4 Outlet valve housing 5 Sealing ring (black) 6 Sleeve 7 Sapphire sleeve The shiny side must point to ruby ball. 8 Ruby ball 9 Ceramic holder for ruby ball 10 Seal The larger opening must point outwards.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 High pressure pump ■ ■ ■ Place the valve housing over the stacked components and hold it tightly. By tilting the tool, the valve components slide into the valve housing. Press the seal by hand well on the valve housing. 6 Checking the flow direction Rinse the valve in the direction of the arrow on the valve housing and check wether liquid is escaping on the other end.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation and maintenance Mounting the pump head Note To prevent the pump head from being positioned the wrong way, it is provided with different bore hole depths for the fastening bolts, i. e. a fastening bolt is longer than all others. The bore hole with the greatest depth must therefore be assigned to the longest bolt. If this is not the case, the pump will not function perfectly. 1 Mount the pump head on the pump again using the four fastening screws (20-5).
■■■■■■■■■■■■■■■■■■■■■■ 6.5 Inline filter Changing the filter The flow must be stopped before changing the filter. 1 Removing the inline filter ■ Unscrew the pressure screws (34-1) from the inline filter. 2 Unscrewing the filter screw ■ Screw the filter screw (34-4) out of the filter housing (34-2) with the aid of two 6.2621.000 adjustable wrenches. 3 Inserting the filter ■ ■ Remove the old filter (34-3) with tweezers. Place the new filter (34-3) flat in the filter housing with tweezers (34-2).
■■■■■■■■■■■■■■■■■■■■■■ 6.6 6 Operation and maintenance Quality Management and validation with Metrohm Quality Management Metrohm offers you comprehensive support in implementing quality management measures for instruments and software. Further information on this can be found in the brochure «Quality Management with Metrohm» available from your local Metrohm agent. Validation Please contact your local Metrohm agent for support in validating instruments and software.
■■■■■■■■■■■■■■■■■■■■■■ 7.1 Problems and their solutions 7 Troubleshooting 7.1 Problems and their solutions Problem Cause Remedy Marked rise in pressure 6.2821.120 inline filter blocked. Replace 6.2821.130 filter (see Chapter 6.5, page 51). Drift of the baseline Eluent – Evaporation of the organic solvent in the eluent. Check the eluent bottle attachment (see Figure 16, page 26). Very noisy baseline High pressure pump – contaminated pump valves. Clean pump valves (see Chapter 6.4.2, page 41).
■■■■■■■■■■■■■■■■■■■■■■ Problem 872 Extension Module IC Pump Cause 7 Troubleshooting Remedy ■ Connect an FEP tubing between the IC instrument and the Extension Module and tighten both clamping screws until they seal.
■■■■■■■■■■■■■■■■■■■■■■ 8.1 Reference conditions 8 Technical specifications 8.1 Reference conditions The technical specifications listed in this chapter refers to the following reference conditions: Ambient temperature +25 °C (±3 °C) Instrument status > 40 minutes in operation (equilibrated) 8.2 Instrument Metal-free IC system IC system Material Painted polyurethane hard foam without CFCs, fire class V0 Intelligent components MagIC Net 8.
■■■■■■■■■■■■■■■■■■■■■■ 8.4 8 Technical specifications Housing Dimensions Width 365 mm Height 131 mm Depth 380 mm Material of base tray, housing and bottle holder 8.5 Polyurethane hard foam (PUR) with flame retardation for fire class UL94V0, CFC-free, coated Eluent degasser Material fluoropolymer Resistance to solvents No restriction (apart from PFC) Build-up time for the vacuum < 60 s 8.
■■■■■■■■■■■■■■■■■■■■■■ 8.7 Interfaces Safety shutdown Function Automatic shutdown upon reaching the pressure limit values Maximum pressure limit ■ Minimum pressure limit ■ ■ ■ ■ ■ Gradient capacity Adjustable from 0.
■■■■■■■■■■■■■■■■■■■■■■ 8.10 8 Technical specifications Weight 1.872.0010 872 Extension Module IC Pump 7.
■■■■■■■■■■■■■■■■■■■■■■ 9.1 Declaration of Conformity 9 Conformity and warranty 9.1 Declaration of Conformity This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company. Name of commodity 872 Extension Module The 872 Extension Module is an expansion tool for upgrading all 850 Professional IC instruments.
■■■■■■■■■■■■■■■■■■■■■■ 9 Conformity and warranty D. Strohm Ch. Buchmann Vice President, Head of R&D Vice President, Head of Production Responsible for Quality Assurance 9.2 Quality Management Principles Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
■■■■■■■■■■■■■■■■■■■■■■ 9.3 Warranty (guarantee) Customer support and service Customer support involves all phases of instrument acquisition and use by the customer, i.e. consulting to define the adequate equipment for the analytical problem at hand, delivery of the equipment, user manuals, training, after-sales service and processing of customer complaints.
■■■■■■■■■■■■■■■■■■■■■■ 10 Accessories 10 Accessories Note Subject to change without notice. 10.1 Scope of delivery 2.872.0010 872 Extension Module – IC Pump Order no. Description 1 1.872.0010 872 Extension Module – IC Pump 1 6.1602.160 Eluent bottle attachment GL 45 Qty. For eluent bottles; with connections for adsorber tube and aspiration tubing. Opening ground joint: A-14/15 1 6.1608.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. 1 Order no. Description 6.1609.000 Adsorber tube / large and bent For filling with adsorber material. Material: Height (mm): Inner diameter (mm): SGJ size: 1 6.1805.060 Glass 129 32 B-14/15 FEP tubing / M6 / 60 cm Protected against light and kink. Material: FEP Inner diameter (mm): 2 Length (mm): 600 1 6.1831.100 PEEK capillary 0.25 mm i.D. / 1 m Material: Outer diameter (inches): Inner diameter (mm): Length (m): 64 ■■■■■■■■ PEEK 1/16 0.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 10 Accessories Order no. Description 6.1834.080 Aspiration tubing, 2 m Aspiration tubing for Professional IC instruments Material: PTFE Outer diameter (mm): 2.5 Inner diameter (mm): 1.5 Length (m): 2 1 6.2023.020 Clip for SGJ 14/15 Clip for SGJ 14/15 Material: 1 6.2156.060 POM Cable extension module - Professional IC, 40 cm Cable for the connection of an extension module to a Professional IC instrument Length (m): 0.4 1 6.2621.030 Length (mm): 1 6.2621.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. 1 Order no. Description 6.2744.070 Pressure screw short Short version. With UNF 10/32 connection. 5 pieces. For the connection of PEEK capillaries Material: PEEK Length (mm): 21 1 6.2744.210 Tubing adapter for aspiration filter For Professional IC instruments 1 6.2758.000 Mixing coil for high presssure gradients For mixing high pressure gradients Material: PEEK 1 6.2816.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 10 Accessories Order no. Description 6.2816.040 Purging needle With PTFE tubing and Luer connection. For syringes. For aspirating eluents. 1 6.2821.090 Aspiration filter Pore size 20 µm. Set of 5 pieces. For 6.1834.000 aspiration tubing and 6.1821.040 and 6.1821.050 filter tubes. Material: PE Outer diameter (mm): 9.5 Length (mm): 35.5 1 6.2821.130 Spare filter for inline filter Spare filter for inline filter. 1 8.872.
10.2 Optional accessories 10.2 ■■■■■■■■■■■■■■■■■■■■■■ Optional accessories 2.872.0010 872 Extension Module – IC Pump Order no. Description 6.2061.100 Flask holder (ProfIC) Flask holder for Professional IC instruments 6.2061.110 Base tray with sensor for Professional IC systems 6.2103.170 Adapter for leak sensor in Professional IC Adapter for the connection of two leak sensors to one Professional IC instrument 6.2156.
■■■■■■■■■■■■■■■■■■■■■■ Order no. Description 6.2744.240 Cross connection UNF 10-32 PEEK 10 Accessories Mixing connection for three solutions 6.2824.130 Macro pump head PEEK Macro pump head for intelligent IC instruments, flow range 0.1...20 mL/min, maximum pressure 12.5 MPa. Material: PEEK (metal-free) 6.5332.000 IC tubing set for a second extension module 6.9988.
■■■■■■■■■■■■■■■■■■■■■■ Index Index Numbers/Symbols 6.2821.090 aspiration filter ...... 40 6.2821.130 filter ...................... 51 Eluent degasser Installation ......................... 28 Technical specifications ....... 57 A Accessories Optional ............................. 68 Scope of delivery ................ 63 Ambient conditions .................. 56 Aspiration filter 6.2821.090 ...... 40 Aspiration tubing for eluent ...... 25 F Filter see also "inline filter" .......... 34 Filter 6.2821.
■■■■■■■■■■■■■■■■■■■■■■ Temperature ............................. 56 Transport ................................. 56 Index Validation ................................. 53 Valves of the high pressure pump ................................................. 49 W Warranty .................................. 62 V Vacuum pump ..........................