Owner's manual

6.5 Inline filter
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
872 Extension Module IC Pump
Changing the filter
The flow must be stopped before changing the filter.
1
Removing the inline filter
Unscrew the pressure screws (34-1) from the inline filter.
2
Unscrewing the filter screw
Screw the filter screw (34-4) out of the filter housing (34-2) with
the aid of two 6.2621.000 adjustable wrenches.
3
Inserting the filter
Remove the old filter (34-3) with tweezers.
Place the new filter (34-3) flat in the filter housing with tweezers
(34-2).
4
Mounting filter screw
Screw the filter screw (34-4) back into the filter housing (34-2)
and tighten by hand. Then additionally tighten slightly with two
6.2621.000 adjustable wrenches.
5
Remounting the inline filter
Screw the pressure screws (34-1) back onto the inline filter.
6
Rinsing the inline filter
Dismantle the guard column (if present) and the separation col-
umn and replace with a 6.2744.040 coupling.
Rinse the instrument with eluent.