Instruction manual

Guidance and manufacturer's declaration - electromagnetic emissions
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Emissions test
RF emissions
EN 55011
RF emissions
EN 55011
Harmonic emissions
EN 61000-3-2
Voltage fluctuations /
flicker emissions
EN 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The Model CP30DW uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause interference in nearby
electronic equipment.
The Model CP30DW is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration - electromagnetic immunity
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical fast
transient/burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
EN 60601 test
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input
/output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
< 5%
U
T
(> 95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
v
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec.
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Model
CP30DW requires continued operation
during power mains interruptions, it is
recommended that the Model CP30DW
be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Compliance
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input
/output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
< 5%
U
T
(> 95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec.
3 A/m
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Compliance with the following EMC guidance is assured only for the use in Europe (220 - 240 V).