Manual

- Press button 1 to validate the selection, the display shows Liagain.
- When pressing button 1 again, the display will show “ FiLt“ (Filter).
- Use button 2 and button 3 to select the desired filter behavior
0 = filter deactivated
1 = filter stage 1: suppresses jumping display values caused be smallest changes
2 = filter stage 2: additional suppression of measuring peaks
(causes delayed reaction of switching output)
- Press button 1 to validate your value, the display shows FiLtagain.
The configuration of the device to the inpu
t signal
is now completed. You now have to configure
the output of the device.
4.2 Selection of the output function
- When pressing button 1 again, the display will show “ outP“ (Output).
- Use button 2 and button 3 (middle or right button) to select the desired output-function.
Description
to select
as output
Output
(out)
see
chapter
no output, device is used as display
no
off
--
2-point-controller
2P
switching
function
5.1
3-point-controller
AL
min-/max-
alarm,
inverse
5.2
- Press button 1 to validate the selected output function. The display shows “ outPagain.
For output function = no the configuration is now finished. Press button 1 to finish the
adjustment and to switch over to display the measuring value.
If the output function was changed here, the delay and preferred state of switching function and
the switching points / alarm-boundaries are to configure in following.
The settings described in the following depend on the output function. Depending on this setting
therefore it is possible that more than one point is present in the following.
- W
hen press
ing button 1 again, the device will display “ 1.dEL“ (delay = delay of switching function).
- Use button 2 and button 3 to set the desired value for the switching-delay.
Hint: The selected value [0.01 ... 2.00] accords the switching delay in seconds.
- Press button 1 to validate the selection. The display shows 1.dELagain.
- When pressing button 1 again, the device will display “ 1.Err
(error = preferred state of switching function).
- Use button 2 and button 3 to set the desired initial state in case of an error.
1010 West Bagley Road, Berea, Ohio 44017 P 440.243.0888 F 440.243.3472 www.noshok.com
9