Outdoor Microphone
Calibration
The design of the M2230-WP Outdoor Measurement Micro-
phone supports easy calibration of the microphone. To cali-
brate, follow the procedure below:
The top section of the M2230-WP is snapped on to •
the body tube. Remove the top section of the M2230-
WP by gently pulling the bird spike upwards. At the
same time gently push up on the cage inside the wind
screen with two fingers of your other hand. You will
feel when the snap mechanism is released.
Gently remove the top section. •
Push
Pull
Calibrate the microphone as described in the XL2 user man-•
ual using the NTi Audio Precision Calibrator.
Snap the top section back into position on the body tube.•
You have successfully calibrated the M2230-WP Outdoor
Measurement Microphone.