User's Manual Part 3

Chapter 5 Capsule Endoscope Procedures
127
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
9. Remove the rubber caps from the real time viewer cable connectors on the
recorder unit (accessible via the opening on the lower right side of the
netted pocket) and on the real time viewer. Using the real time viewer cable,
connect the real time viewer to the recorder unit, as shown in Figure 5.40.
Figure 5.40
If you have mistakenly inserted the recorder unit with its
display panel facing inward, then the recorder unit’s real time
viewer cable connector can be accessed via the opening on
the lower left side of the connector cover.
10. Hold down the real time viewer’s power switch for at least 1 seconds to turn
ON the real time viewer.