User's Manual Part 3

128
Chapter 5 Capsule Endoscope Procedures
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
11. The real time viewer’s display panel will display an icon ( ) indicating
the connection to the recorder unit, and the octagonal display will show “No
RECORD”.
Figure 5.41
When signals from the capsule endoscope are not being
received, or if the reception is poor, “NO RECORD” will be
displayed to the left of the octagonal display, and endoscope
images cannot be recorder by the recorder unit.
On the upper left of the real time viewer’s display panel, a
receiver connection icon is displayed to indicate that the
image is being transmitted from the recorder unit via the real
time viewer cable (see Figure 5.41).
When the real time viewer is receiving images directly from
the capsule endoscope, the direct reception icon is displayed
(see Figure 5.42).
Figure 5.42
Recorder unit connection icon
Direct reception icon