User's Manual Part 3

Chapter 5 Capsule Endoscope Procedures
129
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
5.7 Preparing the capsule endoscope
Do not use a capsule endoscope that have been dropped,
bitten or been subjected to excessive pressure. Using such a
capsule endoscope may result in infection of patient and/or
medical personnel, as well as internal injury to the patient
due to equipment damage.
If the sterile container is open or damaged, the sterility of the
capsule endoscope may have been compromised. Use a
new capsule endoscope instead.
To prevent the capsule endoscope battery from deteriorating,
store the capsule endoscope under a temperature of 0 to
25°C (39 to 77°F). Do not store in a refrigerator or a freezer,
as condensation may result.
Clean the capsule activator with a cotton swab soaked in
70% ethyl or isopropyl alcohol.