Chapter 5 Capsule Endoscope Procedures
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Turning power ON
1.
Slowly remove the sealing paper and take care that the capsule endoscope
does not fall out of the container.
Figure 5.43
2. Hold the sterile container, taking care not to squeeze it. Adjust the direction
of the boss of the capsule activator to the closest groove to the handle of the
inner lid of sterile container.
Then insert the capsule activator straight along with the groove, the capsule
endoscope turn ON (see Figure 5.44). When the capsule endoscope is
turned ON, the LED illuminates.
Figure 5.44
Inner lid of sterile
container
Outer lid of sterile
container
Capsule activator
Insert
Capsule endoscope
insertion hole
Boss
Groove