User's Manual Part 3

Chapter 5 Capsule Endoscope Procedures
137
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Downloading image data
The “Download” function is not available in Endo Capsule
Software Light.
Download the image data from the recorder unit to the workstation according to
the following procedure:
While downloading, do not remove the recorder unit from the
cradle or turn OFF the workstation. Doing so may result in
loss of image data.
While downloading, do not unplug the recorder unit cradle’s
power supply cable. Doing so may result in loss of image
data.
While downloading, do not turn OFF the recorder unit. Doing
so will interrupt the download, and require you to start the
download again from the beginning.
While downloading, do not remove the recorder unit from the
cradle or remove the cradle cable. Doing so may result in
loss of image data. If you do so, connect the recorder to the
workstation through the cradle and the cradle cable and
restart the workstation before downloading again.
If the antenna lead set being worn by a patient is connected
to the recorder unit, do not connect the recorder unit to the
cradle. Otherwise, the patient may experience an electric
shock.
If the workstation cannot recogize the connected recorder(s),
please remove the cradle cable from the workstation, restart
the workstation and then connect it again.
Do not remove the recorder unit from the cradle while
downloading with two recorders connected whether the data
is now set to it or not. Doing so may result in the hangup of
the workstation. If the workstation does not work properly,
please restart the workstation.
1. Check that the recorder unit cradle is connected to the workstation using the
recorder unit cradle cable.
2. Turn ON the workstation and log in, as described in Chapter 4.