User's Manual Part 3

Chapter 5 Capsule Endoscope Procedures
121
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Be sure to use the antenna covers when attaching the
antennas. Failure to do so may prevent the proper reception
of small bowel capsule endoscopic images.
Perform the examination with all 8 antenna pads attached to
the patient. The examination may fail with even one antenna
pad not properly attached.
Do not allow the antenna cable to lay on or near the antenna
pads. It may prevent the proper reception of capsule
endoscopic images.
Do not use an antenna pad that has been completely folded.
Its reception performance may be degraded, causing noise to
appear in the transmitted images.
4. Have the patient get dressed, letting the antenna lead set protrude to the left
of his/her abdomen (see Figure 5.31).
Figure 5.31