Owner's manual

Amplifier 1ch100 User manual 1PA100
October 10, 2004 Release: 1.30 2002 by Pacific Laser Equipment
Manual Operation:
In MC mode the HV output can be controlled manually from 0 up to +150 VDC from the front
panel DC- Offset Potentiometer. In this mode, the HV output is proportional to the number of
turns on the DC offset potentiometer.
1.4 Analog indicator.
Located on the front panel the voltage indicator can be use to read the approximate voltage
applied to the piezo actuator. It is calibrated to swing to the extreme right when the voltage
present at the HV connector reaches 150V.
We recommend that the user should test this indicator in relation with the voltage out and use
the results during the regular operation. I dynamic mode the indicator will average the voltage
output and at high frequency will not be usable.
2. Block diagrams
2.1 General diagram.
PLEquipment, 3941 S. Bristol St. Unit E, # 122, Santa Ana, CA. 92704, 800 984-3482, Fax: 509- 355- 5155, info@plequipment.com
Page 5 of 9