User's Manual

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 119
Performance Tests: Resident Self Tests
Viewing Error Logs
The RST application logs errors during operation of the system and when you run a test. View
the logs by clicking Event Viewer under the Log tab of the RST window and selecting the log
you want to see. Figure 6-12 shows the RST log display.
The logs list errors in chronological order, with the most recent errors at the top of the list. You
can view more detail of an error by double-clicking the entry in the list view.
Figure 6-12 RST Log Display
The Event Viewer includes application logs, a security log, and a system log.
Ultrasound Application Log
This log includes run-time errors encountered by both the ultrasound application and the RST
application. The log lists the date and time the error occurred, the name of the error, the