User's Manual

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 122
Performance Tests: Resident Self Tests
RST Log Codes
The Application Test Log’s Category column contains coded information. The code is in the
following format:
SSnnNN
Where:
SS is the subsystem tested.
BP is the beam processor subsystem.
DP is the display processor subsystem.
nn is the module number of the tested module.
NN is the submodule number.
Example: DP0102
The example indicates that the A/D converter (submodule 02 of module 01) of the display
processor was tested.
To find the meaning of the numeric codes, use the tree control in the RST window’s left side as
described in “Test Modules” on page 117.
Clearing the RST Test Log
1. Right-click RST Test Log in the tree view.
2. In the pop-up menu, select Clear Log.
3. Click Ye s in the confirmation dialog box.