M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 88
Installation: Customer Training
Customer
Training
The customer needs some instruction in maintenance and use of the system:
Instruct the customer how to back up and restore presets (see “Backing Up and Restoring
Presets” on page 143). Make a backup floppy disk of the customer’s presets and configura-
tions, and store it behind the lower enclosure’s left rear door. If the software is reloaded
later, the backup floppy disk lets you quickly restore settings and options.
Explain the importance of letting the system shut down completely before disconnecting the
power cord.
Demonstrate how to clean the air filter (“Cleaning the Air Filter” on page 131), and explain
the importance of the filter to system performance.
Demonstrate how to remove and clean the trackball. (See “Cleaning the Trackball” on
page 132.)
If the system is connected to a DICOM network, show the customer the job queue, and
explain that the queue is empty if all studies have transferred successfully.