Operation Manual

Insulation Test
Insulation resistance is a measure of the integrity of the
insulating materials that separate the electrical windings from
the generator’s steel core. This resistance can degrade over
time or be degraded by contaminants: dust, dirt, oil, grease,
and especially moisture. Most winding failures are due to a
breakdown in the insulation system. In many cases, low
insulation resistance is caused by moisture collected when
the generator is shutdown
Insulation resistance is measured with a megger (mega-ohm
meter). A megger measures insulation resistance by placing
500 volts between the winding and the frame of the
generator. Caution must be taken to remove all electronic
devices (regulators, diodes, surge protectors, capacitors,
protective relays, etc.) from the winding circuit before
checking the insulation. Winding insulation can be checked on
the main stator, main rotor, exciter stator, and exciter rotor.
Minimum resistance is 2 mega-ohms. If the winding
resistance is low it must be dried (see maintenance section)
or repaired.
DIODE TESTING
If the generator is close coupled to an engine, it may be
necessary to "bar over" the engine in order to gain access to
a given area of the rectifier assembly. NEVER use the
generator's fan as a fulcrum to accomplish this. Use the
engine manufacturer's recommended practice to manually
turn over the engine. To prevent possible injury to personnel,
and damage to the equipment, ensure that the engine
cannot start during this procedure.
Remove the two main rotor leads and the three exciter rotor
leads from the rectifier assembly (Figure 5). The rectifier
assembly is now electrically isolated from the generator. The
diodes remain mounted and the diode leads remain
connected to the terminal posts. Using an ohmmeter or a
battery light continuity tester, place one test probe on the
diode lead terminal post. In succession, touch the other test
probe to the lead screw hole in each heat sink. Reverse the
probes and repeat the procedure. You have now tested the
three diodes connected to this terminal post in both the
forward and reverse direction. Repeat the procedure using
the other diode terminal post.
FIGURE 4: DIODE POLARITY
When the positive test probe is connected to the diode's
anode and the negative test probe is connected to the diode's
cathode (forward biased), the diode will switch on and
conduct electricity (Figure 4). This is observed by a low
resistance reading when using an ohm meter or the lighting
of the bulb when using a battery light continuity tester.
Reversing the test leads (reverse biased) will result in the
diode switching off and no electricity will be conducted. The
results of these tests should indicate one of three conditions:
1. Good diode: Will have a much greater resistance in one
direction than the other. Typical reverse biased resistance
will be 30,000 ohms or greater, while forward biased
resistance will be less than 10 ohms. The battery light
tester will have the light "on" in one direction and "off" in
the other.
2. Shorted condition: Ohmmeter reading will be zero, or
very low in both directions. The continuity tester will have
the light "on" in both directions.
3. Open condition: Ohmmeter will have a maximum
(infinity) reading in both directions. Continuity tester light
will be off in both directions.
Diode failure after a 25 hour "run in" period is generally
traceable to external causes such as a lightning strike,
reverse current, line voltage spikes, etc. All 6 diodes are
essentially in the same circuit. When a diode is stressed to
failure, there is no easy method to determine remaining life in
the other diodes. To avoid possible continued failures, it is
recommended that the entire rectifier assembly be replaced
rather than replacing individual diodes.
SERVICE
GENERAL
The service procedures given in this section are those which
can reasonably be conducted on-site with a minimum
number of special tools and equipment. All service
procedures should be conducted by qualified maintenance
personnel. Replacement parts may be ordered through an
authorized service center or directly from the factory.
FIELD FLASHING
Restoring Residual Magnetism
(not applicable on PMG equipped generators)
To restore residual magnetism to the generator, connect a 12
volt battery to the exciter field while the generator using the
following procedure:
1. Shutdown the generator set. Remove the exciter field
leads F+ and F from the regulator.
13
Terminal End
Terminal End
Anode
Anode
(+)
(+)
Cathode
Cathode
(-)
(-)
Stud End
Stud End
Cathode
Cathode
(-)
(-)
Anode
Anode
(+)
(+)
Forward
Forward
Reverse
Reverse