User manual

1. General parameters
Sequence
The name of the sequence.
R-Curve
The oximeter probe manufacturer from the pre-
loaded or user-defined R-curve listing.
2. Sequence parameters
Test #
The test in sequential order.
Ambient light
The ambient light setting, which simulates motion,
ambient sunlight, and 50 Hz or 60 Hz artificial light
interfering with the pulse oximeter’s probe.
Pi
gmentation
The pigmentation setting (low, medium, high)
Preset
Simulated patient conditions. Default is Normal.
SpO
2
[%]
The O
2
saturation level. Range is 30% - 100%.
Rate [BPM]
The pulse rate. Range is 30-250 BPM.
P.Ampl.[%]
The pulse amplitude level. Range is 0% - 100%.
SpO
2
Result
Key in the test result form your Pulse Oximeter
Rate Result
Key in the test result form your Pulse Oximeter
Warning
Key in the test result if the alarm triggered
2. Alarm Limits Test parameters. Information is contained in this
tab is for running alarm limits tests.
Test #
The test in sequential order.
Ambient light
The ambient light setting, which simulates motion,
ambient sunlight, and 50 Hz or 60 Hz artificial light
interfering with the pulse oximeter’s probe.
Pi
gmentation
The pigmentation setting (low, medium, high)
Test Parameter
The value simulated by the QA-510 unit. Select from
Pulse, SpO
2
, Rate, P.Ampl.
ALT Result
Alarm Test Result
Warning
Key in the test result if the alarm triggered
4.2 Sequence Setup
When the Sequence Window is opened the first test parameters for
Make, Amb.L. and Pigm. are set automatically so that, when new
tests are added to the sequence, these parameters are copied auto-
matically to them. To modify or change these parameters, as well as
any other test parameters, click the field and manually enter the value
and press
Enter or select from the drop down list of choices, as ap-
propriate.
1. Single Test Setup:
a. Set the Make, Amb.L. and Pigm. parameters.
b. Select one of the preprogrammed Preset states, or set SpO
2
,
Rate and P.ampl. manually.
4-2