User manual

The Simple mode window contains the following fields and buttons:
1. Sample no. This field is the sample number
2. Perform Simulation. This field contains a checkbox. If the box
is checked the values will be simulated during the set of simula-
tions.
3. Simulated SpO
2
Value. This is the SpO
2
value that you wish to
simulate.
4. SpO
2
Result. This is the field where you key in the value that is
displayed on the SpO
2
meter.
5. Simulate SpO
2
level. This button starts the simulation of the se-
lected SpO
2
values.
6. Calculate R-Curve. This calculates the new R-curve based on
the results from the simulation. The more simulated points that
you use the more accurate the R-curve will be.
7. Transfer R-Curve. This button is used to upload the R-curve
into the memory location in the QA-510 selected in the window
on the right.
When the mouse is over the table, you can press the right mouse but-
ton to insert additional test values or delete extra values.
7-2