17. Press the “Close” Button. This takes us back to our earlier R-
Curve window.
18. Go to “File” and select “Save R-Curve”. This will now save the
new R-Curve to your hard drive.
19. Make sure that the desired upload location is selected in the box
on the right, then press “Upload R-Curve”. The QA-510 will in-
dicate, “Wait” while the data is uploading. Do not disconnect
the cable or turn off the power on the QA-510 during the
download.
20. Turn the power off on the QA-510. Wait 5 sec and turn the
power back on. This refreshes the data in the QA-510.
21. Test the new R-Curve in the QA-510 by selecting the Make that
you uploaded.
22. If the values are off go to step #4 and repeat the procedure now
using this data file as your starting point. Omit steps #7 and #8.
The Advanced mode window contains the following fields and but-
tons:
1. Red Gain value. This field is the gain value for the Red source.
Legal values range from 100-2697.
2. IR Gain value. This field is the gain value for the IR source.
Legal values range from 100-2697.
3. Light Level code. This is the appropriate value selected from
the table in Appendix C. Legal values range from 0-255.
7-4