User manual

APPENDIX B: R-Curves
The following manufacturers’ R-Curves have been pre-loaded into
QA-510, and are provided herein for information.
B1 BCI OXI-Pulse
Make:
BCI OXI-PULSE 3301 71000A1
Reference Unit Serial No.:
320014767
Date:
04.03.97
Red Gain Setting:
2048
IR Gain Setting:
1024
Level Code Setting:
101
Broadband Preamp Gain:
0 dB
IR Preamp Gain:
0 dB
Broadband Trig Threshold:
0.5 V
IR Trig Threshold:
0.5 V
SpO
2
R-value SpO
2
R-value SpO
2
R-value
100% 0.39 78% 1.03 56% 1.57
99% 0.46 77% 1.05 55% 1.60
98% 0.53 76% i .07 54% 1.63
97% 0.59 75% 1.09 53% 1.66
96% 0.63 74% 1.11 52% 1.69
95% 0.66 73% 1.13 51% 1.73
94% 0.69 72% 1.16 50% 1.76
93% 0.71 71% 1.18 49% 1.80
92% 0.74 70% 1.20 46% 1.83
91% 0.76 69% 1.23 47% 1.87
90% 0.78 68% 1.25 46% 1.90
89% 0.80 67% 1.27 45% 1.94
88% 0.83 66% 1.30 44% 1.98
87% 0.85 65% 1.32 43% 2.02
86% 0.87 64% 1.35 42% 2.06
85% 0.89 63% 1.37 41% 2.10
84% 0.91 62% 1.40 40% 2.14
83% 0.93 61% 1.43 39% 2.18
82% 0.95 60% 1.45 38% 2.23
81% 0.97 59% 1.48 37% 2.27
80% 0.99 58% 1.51 36% 2.31
79% 1.01 57% 1.54 35% 2.36
B-1