User manual

B3 Datascope Accutorr
Make:
Datascope Accutorr 3 Sat
Reference Unit Serial No.:
10995F2
Date:
04.03.97
Red Gain Setting:
2048
IR Gain Setting:
1024
Level Code Setting:
110
Broadband Preamp Gain:
0 dB
IR Preamp Gain:
0 dB
Broadband Trig Threshold:
2 V
IR Trig Threshold:
1 V
SpO
2
R-value SpO
2
R-value SpO
2
R-value
100% 0.49 78% 1.22 56% 1.75
99% 0.53 77% 1.25 55% 1.77
98% 0.56 76% 1.28 54% 1.79
97% 0.61 75% 1.31 53% 1.81
96% 0.63 74% 1.33 52% 1.83
95% 0.67 73% 1.36 51% 1.85
94% 0.72 72% 1.38 50% 1.87
93% 0.75 71% 1.41 49% 1.89
92% 0.79 70% 1.43 48% 1.91
91% 0.82 69% 1.46 47% 1.92
90% 0.86 68% 1.48 46% 1.94
89% 0.89 67% 1.51 45% 1.96
88% 0.92 66% 1.53 44% 1.98
87% 0.96 65% 1.55 43% 2.00
86% 0.99 64% 1.58 42% 2.01
85% 1.02 63% 1.60 41% 2.03
84% 1.05 62% 1.62 40% 2.05
83% 1.08 61% 1.64 39% 2.06
82% 1.11 60% 1.67 38% 2.08
81% 1.14 59% 1.69 37% 2.10
80% 1.17 58% 1.71 36% 2.11
79% 1.20 57% 1.73 35% 2.13
B-3