User manual

B6 Marquette Eagle
Make:
Marquette Eagle 3000
Reference Unit Serial No.:
H5LA0215G
Date:
04.03.97
Red Gain Setting:
2048
IR Gain Setting:
1024
Level Code Setting:
101
Broadband Preamp Gain:
0 dB
IR Preamp Gain:
0 dB
Broadband Trig Threshold:
0.5 V
IR Trig Threshold:
0.5 V
SpO
2
R-value SpO
2
R-value SpO
2
R-value
100% 0.50 78% 1.06 56% 1.41
99% 0.53 77% 1.08 55% 1.43
98% 0.57 76% 1.10 54% 1.44
97% 0.60 75% 1.12 53% 1.46
96% 0,63 74% 1,13 52% 1,47
95% 0.66 73% 1.15 51% 1.49
94% 0.69 72% 1.17 50% 1.50
93% 0.72 71% 1.18 49% 1.52
92% 0.75 70% 1.20 46% 1.53
91% 0.77 69% 1.22 47% 1.55
90% 0.80 68% 1.23 46% 1.57
89% 0.83 67% 1.25 45% 1.58
88% 0.85 66% 1.26 44% 1.60
87% 0.88 65% 1.28 43% 1.61
86% 0.90 64% 1.29 42% 1.63
85% 0.92 63% 1.31 41% 1.64
64% 0.94 62% 1.32 40% 1.66
83% 0.96 61% 1.34 39% 1.67
82% 0.99 60% 1.35 38% 1.69
81% 1.01 59% 1.37 37% 1.70
80% 1.03 58% 1.38 36% 1.72
79% 1.04 57% 1.40 35% 1.73