User manual

B7 Nellcor
Make:
Nellcor N-20
Reference Unit Serial No.:
207 74 771
Date:
04.03.97
Red Gain Setting:
2048
IR Gain Setting:
1024
Level Code Setting:
101
Broadband Preamp Gain:
0 dB
IR Preamp Gain:
0 dB
Broadband Trig Threshold:
0.5 V
IR Trig Threshold:
0.5 V
SpO
2
R-value SpO
2
R-value SpO
2
R-value
100% 0.48 78% 1.05 56% 1.40
99% 0.52 77% 1.07 55% 1.42
98% 0.55 76% 1.09 54% 1.43
97% 0.59 75% 1.11 53% 1.45
96% 0.62 74% 1.12 52% 1.46
95% 0.65 73% 1.14 51% 1.48
94% 0.68 72% 1.16 50% 1.49
93% 0.71 71% 1.17 49% 1.51
92% 0.74 70% 1.19 48% 1.52
91% 0.77 69% 1.21 47% 1.54
90% 0.79 68% 1.22 46% 1.55
89% 0.82 67% 1.24 45% 1.57
88% 0.84 68% 1.25 44% 1.59
87% 0.87 65% 1.27 43% 1.60
86% 0.89 64% 1,28 42% 1.62
85% 0.91 63% 1.30 41% 1.63
84% 0.93 62% 1.31 40% 1.65
83% 0.95 61% 1.33 39% 1.66
82% 0.98 60% 1.34 38% 1.68
81% 1.00 59% 1.36 37% 1.69
80% 1.02 58% 1.37 36% 1.71
79% 1.03 57% 1.39 35% 1.72