User manual

B9 Ohmeda OxySTAT
Make:
Ohmeda OxySTAT
Reference Unit Serial No.:
Date:
04.03.97
Red Gain Setting:
480
IR Gain Setting:
640
Level Code Setting:
96
Broadband Preamp Gain:
0 dB
IR Preamp Gain:
0 dB
Broadband Trig Threshold:
0.25 V
IR Trig Threshold:
0.25 V
SpO
2
R-value SpO
2
R-value SpO
2
R-value
100% 0.53 78% 1.26 56% 1.77
99% 0.57 77% 1.29 55% 1.79
98% 0.60 76% 1.31 54% 1.82
97% 0.64 75% 1.34 53% 1.84
96% 0.68 74% 1.36 52% 1.86
95% 0.72 73% 1.39 51% 1.88
94% 0.76 72% 1.41 50% 1.90
93% 0.79 71% 1.43 49% 1.92
92% 0.83 70% 1.46 48% 1.95
91% 0.86 69% 1.48 47% 1.97
90% 0.90 68% 1.50 46% 1.99
89% 0.93 67% 1.53 45% 2.01
88% 0.96 66% 1.55 44% 2.03
87% 1.00 65% 1.57 43% 2.05
86% 1.03 64% 1.59 42% 2.07
85% 1.06 63% 1.62 41% 2.08
84% 1.09 62% 1.64 40% 2.10
83% 1.12 61% 1.66 39% 2.12
82% 1.15 60% 1.68 38% 2.13
81% 1.18 59% 1.70 37% 2.15
80% 1.21 58% 1.73 36% 2.16
79% 1.23 57% 1.75 35% 2.18
B-9