User`s manual

DIGITAL󰚟INCUBATOR
23
CIRCUMSTANCE ITEMS TO BE CONFIRMED SOLUTIONS
Connect the cord again.
Check the main socket with any other
electric appliances.
Try to insert to another outlet.
Check electric cord is properly connected.
Check if there is a power outage.
Check whether the plug is damaged.
Temperature will
not go high enough.
No signs of power
to the incubator.
Check your setting temperature.
Check the air circulation FAN is working.
Initialize the incubator.
[Refer to the page 16]
Set the temperature as required.
Take the power plug out of the outlet, and
disassemble the main body for cleaning
around the air circulation fan with brush.
Supply water.
Set the humidity again as required.
Check your setting humidity.
Close the view window again after checking.
Initialize the incubator. [Refer to the page 16]
Humidity will not
go high enough.
Egg-Turning fails
to work.
Check that the Egg-turning stoppage is not
activated. (Egg-turning lamp lighting)
Check that there are no foreign items on the
egg turning tray.
Check if egg turning plate is assembled correctly.
[Refer to the page 10]
If you need to turn the eggs, you can
manually re-start egg-turning.
Unplug the power supply and
disassemble the main body for cleaning
the egg turning tray and gear.
Assemble egg turning plate as the picture on page 10.
Dew foams in
the incubator.
Check the incubator is not placed
too cold circumstances.
When you turn OFF incubator and then
turn on.
Move the incubator to where the ambient air
temperature is over 20.
Dew condensation is normal.
Dry the incubator completely under well ventilated place.
Take out the plug out of the outlet and
disassemble the main body and clean
around the air circulation fan with brush.
Check if there is any hatching debris such as
feather or eggs-shell in the incubator FAN.
The machine makes
some noise.
- It is normal that the machine
emits some sounds because
it uses air circulation fan
for working.
7. Product Information
(2) Troubleshooting
7. Product Information
(2) Troubleshooting
Check that you have not overfilled the incubator.
Check that the incubator placed on inclined
surface.
Check if it is Dew condensation
Water leaks from
the machine.
Do not fill water over LIMIT indication.
Place incubator under a level surface.
Check if room temperature is too high
Check if incubator is exposed to the direct
rays of sun
Check if room temperature is too low
Check if incubator heater(temperature controlling-
system) is normally working
Adjust room temperature between 20
and 25
.
Install incubator without the direct rays of sun
Return incubator to factory setting [Refer to page 16]
Adjust room temperature between 20
and 25
.
Place thermometer inside of incubator and check
if setting temperature is correct
Return incubator to factory setting [Refer to page 16]
Alarm sounds with
indication
(abnormal high temp.)
Alarm sounds with
indication
(abnormal low temp.)
Hatching result could be affected with various factors, and manufacturer do not have
any responsibility for loss of eggs or birds.
Troubleshooting