User`s manual

24
DIGITAL󰚟INCUBATOR
YES
Parameter󰚟saved
ENTER
SET󰚟UP
BACK
ENTER
* Calibration range : - 9.9 ~ + 9.9
Select with UP/DOWN button
Set with UP/DOWN button
Temp.󰚟CAL.
T e m p . 󰚟 C a l i b r a t i o n
󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 3 5 . 0 ˚ C 󰚟
ESC
SET
Warning message during
pressing ENTER button
Warning message during
pressing ENTER button
WARNING
Fatal Damage with
Incorrect Calibration
Temp.Calibrate
H.CALInfo T.CAL
HOLD
5Sec.
YES
Parameter󰚟saved
ENTER
SET󰚟UP
BACK
ENTER
* Calibration range : -20% ~ +20%
Select with UP/DOWN button
Set with UP/DOWN button
Humi.󰚟CAL.
H u m i . 󰚟 C a l i b r a t i o n
󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 󰚟 3 0 % 󰚟
ESC
SET
WARNING
Fatal Damage with
Incorrect Calibration
HOLD
5Sec.
Humi.Calibrate
H.CALT.CAL
High
Temp
HiErr
Low priced thermometers or hygrometers which you can easily purchase
from the market have large deflection of measurement.
This could cause hatching failure, and it is recommended to use highly
precise thermometer or hygrometer.
Operate incubator for an hour before calibration for temperature stabilization.
And then, observe the temperature or humidity at several position
above 15mm of the egg tray mat with your thermometer, and
calibrate with average value.
As this incubator was made with the optimum conditions set at the factory,
you don't need to calibrate the temperature or humidity before
use except for sensitive birds.
Incubator Section
15mm
Egg Turning -
plate
Thermometer
or Hygrometer
Sensor
7. Product Information
󰚟
(3) How to Calibrate
the Temperature & Humidity
7. Product Information
󰚟
(3) How to Calibrate
the Temperature & Humidity
When MENU screen is shown,
press SET UP/BACK button
and display return to basic screen.
When MENU screen is shown,
press SET UP/BACK button
and display return to basic screen.
<
After Calibration
>
If you can not rely on your thermometer or hygrometer, you can initialize the incubator to factory setting.
(Refer to the page 16)
If you want to raise incubator humidity 5% higher,
you should calibrate 5% lower at humidity
calibration mode.
If you want to raise incubator temperature 0.5 ˚C
higher, you should calibrate 0.5˚C lower at temp.
calibration mode.
Temperature Calibration Humidity Calibration