Note Possible Cause Action
One garment inflates but the second
one does not.
The second garment does not receive
air.
Examine its hoses for kinks, twists and
folds.
The pump stops working, the Status
Indicator turns On in yellow.
Hose is not connected properly to
garment or pump, or prong plug is
not inserted into unused air outlet.
Examine and fasten all air connections.
When treating only one limb, always
plug the unused air outlet with the
prong plug provided with the pump.
If all air connections are OK and the
problem persists, contact Therabody.
The pump works at a very low
pressure, regardless of the pressure
set by the user.
Defective garment Replace garment and check again.
Internal malfunction Contact Therabody.
An irregular noise.
Pump transferring vibrations to a
surface
Make sure the pump is standing even-
ly on all four of its bumpers.
Internal malfunction Contact Therabody.