Safety Instructions

23
Rated Maximum Output
Power of Transmitter (W)
Separation Distance* according
to Frequency of Transmitter (m)
0.01 0.23
0.1 0.73
1 2.3
10 7.3
100 23
EMC Manufacturer Declarations
Electromagnetic Compatibility (EMC) Statement for Home Healthcare Environment
The RecoveryAir PRO system has been evaluated to international standard IEC 60601- 1-2 “General requirements for
basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements
and tests”.
The Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided here within the accompanying documents.
Portable and mobile RF communication can affect the Medical Electrical Equipment. See below recommended
separation distances between portable and mobile RF communication equipment and the RecoveryAir PRO system.
Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones
and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d = 3.3 m away
from the system.
*Note: The distance calculated from 800 MHz to 2.5 GHz