User`s guide

10 230750-101, Rev. C.
TONO-PEN AVIA VET VERIFICATION
If suspect readings are observed, a Verication Procedure routine
should be performed. This will ensure that the sensor and elec-
tronics are performing correctly.
VERIFICATION PROCEDURE
Hold the TONO-PEN AVIA VET tonometer with the sensor
end pointing down towards the oor.
Press and hold the activation button for 5 seconds -
a beep will sound at one second intervals.
At the end of the 5 second button hold, the display will
show “dn”.
Keep the pen vertical, with the sensor tip pointing down
towards the oor, for a total of 15 seconds.
At the end of this period, a beep will sound and the display
will show “UP
Immediately point the sensor tip straight up and wait for
the next beep (within 3 seconds).
A properly functioning TONO-PEN AVIA VET tonometer
will display “Pass. Pressing the activation button will now
put the device into Applanation Mode.
This complete test will need to be repeated if the display
shows “Fail. A single press of the Activation button will
re-start the Verication Procedure.
Note: the TONO-PEN AVIA VET tonometer will
remain in the Verication mode until it passes the
Verication Procedure.
The most common cause for failure is debris around the
sensor tip. Follow the instructions for cleaning the tip in
the Maintenance & Storage section before re-performing
the Verication Procedure.
If this is unsuccessful, please contact the Reichert
Technical Service Department.