User`s guide
16 230750-101, Rev. C.
Verication Mode
A Verication Procedure may be performed to ensure that the
sensor and electronics are performing as expected. A Verication
Procedure may also be initiated if suspect readings are observed.
(See earlier section TONO-PEN AVIA VET Verication for details.)
If the Verication Procedure resulted with the LCD displaying
“Fail”, repeat the Verication Procedure, or reference the
Troubleshooting section for potential causes for test failure.
Note: The TONO-PEN AVIA VET tonometer will remain in the
Verication mode until it passes the Verication Procedure.