User`s guide

18 230750-101, Rev. C.
Applanation
Press and release the activation button once to initiate the 1.
applanation mode.
A brief display of “882. 888” will ash on the LCD, a double row of
dashes “===” will be displayed, the green LED will turn on, and
a “beep” tone will sound, indicating the TONO-PEN AVIA VET
tonometer is ready to measure intraocular pressure.
Tap the sensor very lightly and briey on the corneal surface, the 3.
device will chirp and the DATA eld on the LCD will increment for
each valid IOP reading obtained.
When another beep tone is heard, indicating 6 applanations 4.
have been read, the green LED will turn OFF, and the averaged
IOP measurement will appear on the LCD above “mmHg”. The
statistical condence indicator will appear on the LCD above
DATA.
Notes:
After pressing the Activation button, do not shake the
TONO-PEN AVIA VET tonometer as it may register a reading.
If this occurs simply re-press the Activation button and start
the measurement again.
If 15 seconds elapse prior to applanation being initiated, the
TONO-PEN AVIA VET tonometer will initiate the Sleep Mode,
indicated initially by the display of a single row of dashes
“---”, followed by a blank display. The applanation mode may be
initiated by pressing the activation button.
If any error codes appear on the LCD after the nal beep, the
applanation procedure must be repeated.