User`s guide

24 230750-101, Rev. C.
Guidance & Manufacturers Declaration
Table 201 – Guidance and Manufacturers Declaration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The
TONO-PEN AVIA VET is intended for use in the electromagnetic environment specied
below. The customer or user of the
TONO-PEN AVIA VET should ensure that it is used in such
an environment.
Emissions Test Compliance
Electromagnetic Environment
- Guidance -
RF Emissions
CISPR 11
Group 1
Class B
The
TONO-PEN AVIA VET uses RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
Harmonics
IEC 61000-3-2
N/A
The
TONO-PEN AVIA VET is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies building for
domestic power.
Flicker
IEC 61000-3-3
N/A
Table 202 – Guidance and Manufacturers Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
TONO-PEN AVIA VET is suitable for use in electromagnetic environment specied below.
The customer or user of the
TONO-PEN AVIA VET should ensure that it is used in such an
environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete or ceramic
tile. If oors are synthetic, the R/H should be
at least 30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
NA
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
NA
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
NA
Mains power quality should be that of a typical
residential, commercial or hospital environ-
ment. If the user of the
TONO-PEN AVIA
VET
requires continued operation during
power mains interruptions, it is recommended
that the
TONO-PEN AVIA VET be powered
from an uninterruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic elds should be that
of a typical residential, commercial or hospital
environment.