Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 53/90
3.9.2 Radiographic operating
1) General radiography method
1. Fit the wide X-ray Field into the Table Center
2. Move a patient to the Table.]
3. Insert the Cassette into the part that you will take X-ray
(When you take X-ray directly on the affected part...)
4. If you take X-ray with using Bucky, insert the Cassette after pulling the Cassette Tray out.
5. After putting the Cassette, fix it with lock device, not to make movement.
(Push the Cassette Tray into the Table inside, then it is fixed by the lock device.)
6. Move directly toward the part which you want to take X-
ray, using Table Sliding Button or
tube.
7. Check the position and check again the Cassette position.
(When you need to adjust again, repeat above 5.)
8. Explain the suggestions about breathing and movement to a patient.
9. X-ray exposure control
For the Hand -operated Mode
Collate the tube voltage, tube current and time for taking X-ray
For the AEC Mode
1) Select the position for the Field Selection
2)
You can select the singular or plural position for the Field
Selection.
In the AEC Mode, if the X-ray quantity is accumulated adequately, time is controled
automatically to generate the regular picture strength regardless of the subject's thickness.
After turning off the LED by pressing the READY button, finish taking
X-ray by pressing the Exposure button.
Check the follow point, when you finish READY.
Check whether Table Bucky operates well or not, when you select BUCKY
1(Check the sound)
Before putting
The cassette
Afer putting
The Cassette,
make under lock