Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 55/90
8. Control the fluoroscopic tube voltage and tube current.
In the manual mode
According to the patient's physical condition, control the fluoroscopic tube voltage and tube
current through the monitor.
REX-RF fluoroscopy console
In the ABC mode
Set the ABC mode by pressing the ABC function key at the console.
In the ABC mode, fluoroscopic tube voltage and tube current are changed adequately
according to the X-ray absorption of a subject, and they control the monitor
brightness regularly.
9. Begin to scan for catching the adequate part which you'll take X-ray, moving the Table and
Tower part.
Inclination & Sliding Rele
ase all lock Fixation for up/down
Inverse Inclination Movement of Tower movement of Tower
10. If the scan image is catched adequately, you can control the fluoroscopic iris to describe the
detail region and clearer anatomical image. Control the range of the X-ray Field, then you can
get clearer image due to the reduction of scattering line.
If you move the iris screw to left, it becomes narrower with right/left same width.
If you move the iris screw to right, it becomes wider with right/left same width.
If you move the iris screw to up, it becomes wider with up/down same width.
If you move the iris screw to down, it becomes narrower with up/down same width.
11. If you complete the preparation for taking X-ray with SPOT, lock the Tower to prevent it from
moving.
All lock Fixation for up/down
the Tower movement of Tower