Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 61/90
5. After putting the Cassette, fix it with lock device, not to make movement.
(Push the Cassette Tray into the Table inside, then it is fixed by the lock device.)
6. Move the Table Board to the part which you want to take X-ray, using Table Sensor.
7. Check the position and check again the Cassette position.
(When you need to adjust again, repeat above 5.)
8. Explain the suggestions about breathing and movement to a patient.
9. X-ray Irradiation Control
For the Hand -operated Mode
Collate adequately the tube voltage, tube current and time for taking X-ray
For the AEC Mode
1) Select the position for the Field Selection
2) You can select the singular or plural position for the Field Selection.
In the AEC Mode, if the X-ray quantity is accumulated a
dequately, time is controled
automatically to generate the regular picture strength regardless of the subject's thickness.
10. Finish Taking X-ray
- After taking X-ray, relocate the Table position originally.
After turning off the LED by pressing the READY button, finish
taking X-ray by pressing the EXPOSURE button. Check the follow
point, when you finish READY.
Check whether Table Bucky operates well or not, when you select
BUCKY 1 (Check the sound)
Afer putting
The Cassette,
make under
lock