Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 68/90
Please, refer to[3.3 power off] more detail.
2. Perform cleaning and sterilization fluid.
3. After completion, please check followings before supplying power.
Dry water or sterilization fluid completely.
Arrange tools that are used for cleaning and sterilization.
4.1.3 Periodical Inspection
In order to maintain the system in best condition, it is important to have periodic maintenance and
inspection schedule before damages or errors.
Below maintenance schedule shows the minimum frequency. This is the minimum schedule that
has to be performed and if necessary, additional maintenances have to be performed.
Record all performed maintenance services and changed data on Data Book.
Check cables
Check for correct connections and proper strappings for all the wires. Wires cannot be
exposed from connectors and check for proper and complete cable covers.
In case of errors or damages on the system check for basic problems including power
supply or fuse problems. Service technician from the manufacturer must perform
other services or repairs.
1) Connection ground cable
The centeral standard grounding connection of the system is located in Power Cabinet.
Check minimum resistance range of ground lead wires with multi-meter.