Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 69/90
In case of errors or damages on the system check for basic problems including power
supply or fuse problems. Service technician from the manufacturer must perform
other services or repairs.
2) AC power supply for exposure room
Check AC power supply value between the phase, neutral, and grounding wires. These
values must be within tolerance level of initial installation.
In case of errors or damages on the system check for basic problems including power
supply or fuse problems. Service technician from the manufacturer must perform
other services or repairs.
daily
- Operational condition of detector power device
- Operational condition of detector
- Operational condition or errors on system computer
- Cleanness of detector device
- Examine test
- Drive test andexamine focused test
weekly
- Tightening condition of high voltage terminal
- Wiring condition.
- Screws tightening condition
- Cleanness of Generator
- Check surface conditions of detector, grid and ion chamber
monthly
- Generator oiling condition.
- Image Resolution Test.
- Safety stopper fixation condition.
- Gap adjustment and operating conditions of Locks
- Check wire rope
quarter
- X –Ray examination condition
- Deformation status of installed parts
- Difference between collimation and actual X-Ray collimator
- Wiring condition