Owner`s manual

Operator manual REX-***R(F)
LISTEM Corporation 75/90
5.2 Highvoltage generator
1) REX-RF
MODEL REX-850RF REX-650RF REX-550RF REX-525RF
Phase
Voltage
(±10%)
400VAC 400VAC 400VAC 400VAC
Frequency 50/60Hz 50/60Hz 50/60Hz 50/60Hz
System
Power Line
Requirement
Power Input
KVA
100KVA 80KVA 62.5KVA 50KVA
Power capacity 80KW 64kW 50kW 40kW
Maximum mA at
Maximum KVP
800/100 400/150 320/150 320/125
Rating
(Rad)
Maximum KVP at
Maximum mA
500/150 100/640 100/500 80/500
Syste
m
Rating
(Fluor)
Maximum mA at
Maximum KVP
4mA/120kVp 4mA/120kVp 4mA/120kVp 4mA/120kVp
Maximum mA 800 640 500 500
Fluoro Maximum mA 4mA (Higher mA-upto 6mA- : Option)
Maximum KVP 150 150 150 125
Fluoro Maximum KVP 120kVp
mA Range
(±20% precision)
10~800,
(20steps)
10~600,
(19Steps)
10~500, (18Steps)
Fluoro mA Range
(±20% precision)
0.4~4mA (19Steps) (Higher mA – upto 6mA : Optional)
mAs Range 0.01~600
KVP Range
( ±10% precision )
40~150kVP, (1KVp Step)
40~125kVp,
(1KVP Step)
Fluoro KVP Range
( ±10% precision )
40~120kVp, (1kVp Step)
Time Range
( ±10% +1ms precision )
1ms~10sec (38 Steps)
Fluoro Time Range 0~10Minutes (available by user’s selection)
Exposure Indication Audible and Visible
mA, kVp selector Push Button
Self Diagnostic System Yes
100mA Small 100mA Small 100mA Small 100mA Small
Anode Current MAX.
800mA Large 640mA Large 500mA Large 500mA Large