User Manual - Version 2.2 Manual

Counterpart User Manual - Configuration 45
3.7 Calibration Menu
See Section 4.0 on page 49 for calibration procedures. The Calibration menu can be protected by assigning a
password in the Feature menu.
The Counterpart requires the WZERO and WSPAN points to be calibrated. The linearity points are
optional, but must NOT duplicate zero or span. During calibration, the TARE key acts as a data entry
confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.
Figure 3-13. Calibration Menu
Table 3-9. Calibration Menu Parameters
CALIBR Menu
Parameter Choices Description
Level 2 submenus
WZERO Press the calibrate softkey and the A/D
raw counts will be displayed.
Press calibrate softkey again to calibrate zero, or press MENU to cancel. “CALIBRATING” will
be appear pr
ior to automatically moving to WVAL.
WVAL Press calibrate softkey to display and edit the test weight value. Press calibrate softkey again
to move to WSPAN.
WSPAN Press calibrate softkey and the A/D raw counts will be displayed. Pr
ess calibrate softkey to
calibrate the span or press MENU to cancel. “CALIBRATING” will be appear prior to
automatically moving to WLIN.
WLIN POINT 1 —
POI
NT 5
WGT 1 allows you to display and edit the test weight value. Press calibrate softkey to edit the
value. CAL1 allows you to calibrate and display the raw A/D value. Press TARE to perform
calibration. Press calibrate softkey to will move to the next calibration point.
The linearity points are optional, but must NOT duplicate zero or span.
They must be between zero and span.
REZERO Press calibrate softkey to remove an offset value
from the zero and span calibrations (e.g. if
hooks or chains are being used during calibration).
Use this parameter only after WZERO and WSPAN have been set. See
Section 4.1 on page 49 for more information about using this parameter.
Note
WZERO
Calibrating
Calibrating
Display and edi
Press the calibrate
softkey to calibrate
zero.
t
test weight value
WVAL WSPAN
Press calibrate
softkey to remove
offset from zero from
zero and span
calibrations
REZEROWLIN
POINT 1
POINT 2 POINT 3 POINT 4 POINT 5
Same as POINT 1
WGT 1
CAL1
Calibrating
Press the calibrate
softkey to calibrate
span.
XXXXXXX
ID
DISPLAYTESTSETUPAUDIT CALIBR
Press Menu
to cancel
Press Menu to
cancel
Previous A/D raw
counts are shown
Previous A/D raw
counts are shown
SCALE 1 SCALE 2
SETPTS
Note
Note