User Manual - Version 2.2 Manual

Counterpart User Manual - Serial Commands 65
6.1.5 Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 6-4 through 6-12).
Current configuration parameter settings can be
displayed
in all modes using the following syntax:
command<ENTER>
Most parameter values can be changed in menu mode
only.
Use the following command syntax when changing
parameter values:
command=value<ENTER>, where value
is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads
??.
For example, to set the motion band parameter to 5
divisions, type the
following:
SC.MOTBAND=5<ENTER>
For parameters with selectable values, enter the
command and equal sign followed by a question
mark:
command=?<ENTER> to see a list of those values.
The indicator must be in menu mode to use this
function.
Some parameters are valid only if other
parameters or parameter values are
specified. Rest
rictions for front-panel
configuration also apply to serial command configuration.
Note
Command Description Values
SC.GRADS#n Graduations 1–100000
SC.ZTRKBND#n Zero track band 0, 0–100
SC.ZRANGE#n Zero range 1.900000, 0–100
SC.MOTBAND#n Motion band 1, 0–100
SC.SSTIME#n Standstill time 1–65535
SC.OVRLOAD#n Overload FS+2%, FS+1D, FS+9D, FS
SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n
Digital filtering 1, 2, 4, 8, 16, 32, 64, 128, 256
SC.DFSENS#n Digital filter cutout sensitivity 2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT,
12
8OUT
SC.DFTHRH#n Digital filter cutout threshold NONE, 2D, 5D, 10D, 20D, 50D, 100D, 200D,
25
0D
SC.THRESH#n Zero threshold 099
99999
SC.SMPRAT#n Sample rate 30HZ, 60HZ, 120HZ, 240HZ, 480HZ, 960HZ
SC.PWRUPMD#n Power up mode GO, DELAY
SC.TAREFN#n Tare function BOTH, NOTARE, PBTARE, KEYED
SC.PRI.DECPNT#n Primary units decimal position 8.888888, 88.88888, 888.8888, 8888.888,
88
888.88, 888888.8, 8888888, 8888880,
8888800
SC.PRI.DSPDIV#n Primary units display divisions 1D, 2D, 5D
SC.PRI.UNITS#n Primary units lb, kg, g, oz, OFF
SC.ACCUM#n Accumulator enable ON, OFF
SC.WZERO#n Zero calibration
SC.WVAL#n Test weight value test_weight_value
SC.WSPAN#n Span calibration
SC.WLIN.F1–
SC.WLIN.F5#n
Actual raw count value for linearization
points
1–5
0–16777215
Table 6-4. SCALES Serial Commands