User manual

RIGOL Chapter 5 Specifications
5-2 DP800A Users Guide
Rise
Full Load <11ms <50ms
No Load
<10ms
<25ms
Fall
Full Load <13ms <30ms
No Load <200ms <400ms
OVP/OCP
Accuracy ±(Output
Percentage+Offset)
0.5%+0.5V/0.5%+0.5A
Activation Time
1.5ms (OVP≥ 3V)
<10ms (OVP<3V and OCP)
Mechanical
Dimensions 239mm(W) x 157mm(H) x 418mm(D)
Weight 9.0kg
Power
AC Input
(50Hz-60Hz)
100Vac+10%, 115Vac+10%,230Vac+10% (maximum 250VAC)
I/O
USB Device
1
USB Host 1
LAN 1
RS232
1
Digital IO 1
Environment
Working Temperature
Full Rated Value Output: 0
to 40
Under Relatively Higher Temperature: the linearity of the output
current reduces to 50% at the highest temperature 55
Cooling Method
Fan Cooling
Note:
[1] The accuracy parameters are acquired via calibration under 25 after 1-hour warm-up.
[2] The maximum time required for the output to change accordingly after receiving the APPLy
and SOURce commands.
[3] The variation of the output within 8 hours after 30-minute warm-up when the load circuit and
environment temperature are constant.