RIGOL Chapter 2 Front Panel Operations
2-14 DP800A Users Guide
3. Edit the templet parameters
After selecting the desired templet, set the templet parameters. For different
templets, the parameters to be set are different as shown in Table 2-1.
Table 2-1 Templet Parameters
Templet
Parameter
Sine
Max Value, Min Value, Points, Interval, Inverted
Pulse Hi Level, Lo Level, Width, Period, Inverted
Ramp Max Value, Min Value, Points, Interval, Symmetry, Inverted
Stair Up Max Value, Min Value, Points, Interval
Stair Dn Max Value, Min Value, Points, Interval
StairUpDn
Max Value, Min Value, Points, Interval
Exp Rise
Max Value, Min Value, Points, Interval, Rise Index
Exp Fall Max Value, Min Value, Points, Interval, Fall Index
Max Value
Set the maximum voltage or current of the templet currently selected. The
range is related to the channel currently selected.
Min Value
Set the minimum voltage or current of the templet currently selected. The
range is related to the channel currently selected. The minimum cannot be
greater than the maximum currently set.
Points
The total number of points refers to the number of groups of timer
parameters created using the templet currently selected. The range is from
10 to 1048.
When the total number of points (denoted by P) and the current number of
output groups (denoted by G) are different, P groups of parameters will be
created using the templet and then, the number of output groups will
change to P automatically.
Interval
The interval refers to the time required for the instrument to output each
group of timer parameters created using the templet currently selected and
the range is from 1s to 99999s.
Inverted
When the templet currently selected is Sine, Pulse or Ramp, if Inverted is
enabled, the instrument will first turn the preset waveform upside down
and then create timer parameters.