Operating instructions

INSTrument Subsystem R&S ESCI
1166.6004.12 6.114 E-1
INSTrument:COUPle:ATTenuation ALL | NONE
This command couples the RF attenuation settings for the analyzer and receiver modes to each
other. The RF attenuation and level unit are set to the same value in the analyzer and receiver
screens.
Example: "INST:COUP:ATT ALL"
Characteristics: *RST value: ALL
SCPI: device specific
Mode: R, A
INSTrument:COUPle:BANDwidth|BWIDth ALL | NONE
This command couples the bandwidth settings for the analyzer and receiver modes to each other.
The resolution bandwidth (RBW) and filter type are set to the same value in the analyzer and
receiver screens. Other settings, such as the quasipeak detector, may prevent these settings.
Example: "INST:COUP:BWID ALL"
Characteristics: *RST value: NONE
SCPI: device specific
Mode: R, A
INSTrument:COUPle:CENTer ALL | NONE
This command couples the center frequency or receive frequency settings for the analyzer and
receiver modes to each other.
Example: "INST:COUP:CENT ALL"
Characteristics: *RST value: ALL
SCPI: device specific
Mode: R, A
INSTrument:COUPle:DEModulation ALL | NONE
This command couples the audio demodulator settings for the analyzer and receiver modes to each
other.
Example: "INST:COUP:DEM ALL"
Characteristics: *RST value: NONE
SCPI: device specific
Mode: R, A
INSTrument:COUPle:GAIN ALL | NONE
This command couples the preamplifier settings for the analyzer and receiver modes to each other.
Example: "INST:COUP:GAIN ALL"
Characteristics: *RST value: ALL
SCPI: device specific
Mode: R, A