Operating instructions

MMEMory Subsystem R&S ESCI
1166.6004.12 6.124 E-1
MMEMory:SELect[:ITEM]:LINes:ALL ON | OFF
This command adds all limit lines (activated and de-activated) to the list of device settings to be stored/loaded.
Example: "MMEM:SEL:LIN:ALL ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: all
MMEMory:SELect[:ITEM]:FINal ON | OFF
This command includes the final measurement data or the peak list in the list of partial datasets of a
device setting to be stored/loaded.
Example: "MMEM:SEL:FIN ON" 'inserts the final measurement data in the list of partial
'data sets
Characteristics: *RST value: ON
SCPI: device-specific
Modes: all
MMEMory:SELect[:ITEM]:SCData ON | OFF
This command adds the tracking generator calibration data to the list of device settings to be stored/loaded.
Example: "MMEM:SEL:SCD ON" 'Inserts the tracking generator correction data
'in the list of data subsets
Characteristics: *RST value: OFF
SCPI: device-specific
Modes: all
This command is only available in conjunction with the tracking generator option R&S FSP-B9 or
external generator control option R&S FSP-B10.
MMEMory:SELect[:ITEM]:TRANsducer:ALL ON | OFF
This command includes all transducer factors and sets in the list of data subsets of a device setting
to be stored/loaded.
Example: "MMEM:SEL:TRAN:ALL ON"
Characteristics: *RST value: ON
SCPI: device-specific
Modes: all
MMEMory:SELect[:ITEM]:ALL
This command includes all data subsets in the list device settings to be stored/loaded.
Example: "MMEM:SEL:ALL"
Characteristics: *RST value: --
SCPI: device-specific
Mode: all