Operating instructions
R&S ESCI Status Reporting System
1166.6004.12 5.33 E-1
Resetting Values of the Status Reporting System
Table 5-12 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST and SYSTem:PRESet influences the functional
instrument settings. In particular, DCL does not change the instrument settings.
Table 5-12 Resettting instrument functions
Event Switching on supply
voltage DCL,SDC
Power-On-Status-
Clear
(Device Clear,
Selected Device
Clear)
*RST or
SYSTem:PRESet
STATus:PRESet *CLS
Effect 0 1
Clear STB,ESR
yes
yes
Clear SRE,ESE
yes
Clear PPE
yes
Clear EVENTt parts of the
registers
yes
yes
Clear Enable parts of all
OPERation and
QUEStionable registers,
Fill Enable parts of all
other registers with "1".
yes
yes
Fill PTRansition parts with
"1" ,
Clear NTRansition parts
yes
yes
Clear error queue yes yes
yes
Clear output buffer yes yes yes 1) 1) 1)
Clear command
processing and input
buffer
yes yes yes
1) Every command being the first in a command line, i.e., immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer.