Technical data

Manual
Double Sheet Detector R1000 series E20
Start-up
B0046991 / Rev. 1.6
100
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
7.2 Operation
Note The system is factory adjusted for “single measurement“. Therefore the
system functions only if the signal “measurement start“ is activated. The
modification of the configuration is described in the following sections.
After applying power to the system, the system type and the version number are briefly displayed.
Afterwards the system switches into that measurement menu which was active before the unit was
switched off.
1 LED functions
2 Pushbutton functions
3 Program number
4 Selected sensor
5 Type of material
6 Nominal thickness
7 Measurement status
8 Lower limit value
9 Sensor number with measured value
10 Upper limit value
Fig. 36: Frontal view of the R1000 with enlarged display detail
LED functions (symbols above the keys)
<
ON, if the measured value reaches (or falls short of) the lower limit value.
Norm
ON, if the measured value is within the lower and upper limit values.
>
ON, if the measured value reaches (or exceeds) the upper limit value.
Enable
- blinks, if a fault is present
- ON, if the equipment is in measurement mode
- OFF, during parameter programming with the keyboard
Key functions (symbols below the keys)
Changes the parameters (forward)
Changes the parameters (backwards)
Enter
Selects a parameter for changing and confirming the change
Menu
Activates the menu level, for deleting a fault message and exiting the menu level.