Technical data

Manual
Double Sheet Detector R1000 series E20
B0046991 / Rev. 1.6
Start-up
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
105
7.6 Program parameters
The program parameters contain the information required for the various measurement programs.
For accessing the program parameters, select the mode “configuration program parameter“ and
confirm with ENTER.
Program parameter 1: Selecting the measuring program to be configured.
Fig. 43: Selecting the program to be configured
Here the measuring program (1-255) to be configured is selected. All following adjustments apply
to the selected program (here: program no. 1).
Program parameter 2: Selecting the sensor to be used for measurement (only for 4P version).
Fig. 44: Selecting the sensor
Here the sensor (1 - 4) must be selected, which is to be used for measurement (4P version only).
For this purpose the item "program" must be selected under “sensor selection“ in the system
configuration.
If the sensor selection is made via "program" the sequence measurements are also available.
Thus the following combinations are available:
• 1 / 2 / 3 / 4 / 1+2 / 1+3 / 1+4 / 2+3 / 2+4 / 3+4 / 1+2+3 / 1+2+4 / 1+3+4 / 2+3+4 / 1+2+3+4
Program parameter 3: Specifying the nominal thickness of the sheet to be measured.
Fig. 45: Adjusting the nominal thickness
Here the nominal thickness of the sheet to be measured is specified. Normally, the nominal
thickness is the sheet thickness used for teach-in.
Note A change of the nominal thickness deletes the values determined
during teach-in. A new teach-in is required.
program par. 1
program no.: 1
program par. 2
sensor: 1
program par. 3
nom.thick.0.90mm