Perform a device reset
7.2.8 Simulation active – Check operational mode
Alarm classification
Device status
Functional Check
Class
Warning
Possible cause
The device is in simulation mode and is not reporting actual information.
Recommended actions
1. If this behavior is not desired, stop simulation mode.
2.
If the condition persists, restart the device.
Related information
Use the simulation mode
Perform a device reset
7.2.9 Device temperature over-run – Clear source of heat
Alarm classification
Device status
Out of specification
Class
Warning
Possible cause
The electronics temperature is outside the operating range.
Recommended actions
1. Verify ambient temperature is within the operating range.
2.
Clear source of heat.
Related information
Ambient temperature
7.2.10 Device temperature under-run – Insulate device
Alarm classification
Device status
Out of specification
Class
Warning
Possible cause
The electronics temperature is outside the operating range.
Reference Manual Service and troubleshooting